65
5. System settings
5-1. Lock and language settings
Touch the
button on the HOME Screen to show the
SETTINGS screen.
LOCK STATUS
The calibration (CAL), work (WORK) and all (ALL) lock states
are shown with icons.
Locked
Unlocked
Touch an icon to open the lock status setting screen.
o Password locking can be set for the ALL LOCK option.
The password is fixed to “6803”.
LANGUAGE
Use to set the display language.
5-1-1. Lock settings
Changing calibration values and work setting values can be
prohibited.
Touch a lock status icon on the SETTINGS screen to change the
lock status.
To change a setting, touch the setting value display button
and select UNLOCK or LOCK.
o The calibration value lock setting can also be changed on
the load cell and displacement sensor screens.
They both change the same calibration value lock setting.
Locking the calibration value for only load cells or only
displacement sensors is not possible.
o Values for locked settings appear green in the menu and
cannot be changed.
o ALL LOCK prevents opening any SETTINGS menu other
than LOCK STATUS. This also activates CALIBRATION LOCK
and WORK LOCK.
Deactivating ALL LOCK also deactivates CALIBRATION LOCK
and WORK LOCK.
5-1-2. Language setting
This can be set to Japanese, English, Chinese or Korean.