4-5-2. BAND JUDGMENT
Measurement waveforms are compared to comparison wave-
forms to conduct judgment.
Instead of high limit, low limit and other fixed values for
continuous judgment, any curved line can be used for
comparison.
Band judgment is conducted when “REC is the status dis-
played on the screen.
High limit
value
Low limit
value
¦qStart judgment
4-5-2-1. Band judgment settings
Touch the buttons in the following order on the Home Screen.
e
WORK
e
BAND JUDGMENT
This sets management values for the load and judges whether
these values are exceeded.
To use band judgment, after setting SELECT REF. WAVE to SELECT,
set WAVEFORM COMP. to ENABLE.
o If no reference waveform has been set, the following
WAVEFORM COMP. item cannot be set to “ENABLE”.
4. Settings
SELECT REF. WAVE
Set the waveform to use for reference.
WAVE SAMPLING
Use measured data for reference.
WAVE LOADING
Load a waveform saved in the unit.