EasyManuals Logo
Home>Tektronix>Test Equipment>2230

Tektronix 2230 Service Manual

Tektronix 2230
483 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #156 background imageLoading...
Page #156 background image
Adjustment Procedure 2230 Service
h. ADJUSTMF/LF Comp (C3) and MF/LF Gain Bal
(R47) for the best front corner and flat top.
i. Disconnect the test equipment from the instrument.
6. Adjust Vertical Gain (R145, R195, R76, and R26)
a. Connect a 50 mV standard-amplitude signal from
the Calibration Generator via a 50 0 cable to the CH 1 OR
X input connector.
b. Set the A SEC/DIV switch to 0.2 ms.
c. Center the display within the graticule using the
Channel 1 POSITION control.
d. ADJUSTCh 1 Gain (R145) for an exact 5-division
display.
e. Move the cable from the CH 1 OR X input connector
to the CH 2 OR Y input connector. Set the VERTICAL
MODE switch to CH 2.
f. Center the display within the graticule using the
Channel 2 POSITION control.
g. ADJUSTCh 2 Gain (R195) for an exact 5-division
display.
h. Change the generator output to 10 mV and set both
VOLTS/DIV switches to 2 mV.
i. Repeat parts d and g until the gain of the two chan
nels is identical.
J. ADJUST2 mV Gain (R76) for an exact 5-division
display.
k. Move the cable from the CH 2 OR Y input connector
to the CH 1 OR X input connector. Set the VERTICAL
MODE switch to CH 1.
l. ADJUST2 mV Gain (R26) for an exact 5-division
display.
m. Set both AC-GND-DC switches to GND.
n. CHECKThat no trace shift occurs when switching
between the 5 mV and 2 mV positions of the CH 1
VOLTS/DIV switch. If trace shift is observed, repeat
Step 2 of this procedure.
o. Set the VERTICAL MODE switch to CH 2.
p. CHECK—That no trace shift occurs when switching
between the 5 mV and 2 mV positions of the CH 2
VOLTS/DIV switch. If trace shift is observed, repeat
Step 2 of this procedure.
7. Check Deflection Accuracy and Variable Range
a. Set:
VERTICAL MODE CH 1
AC-GND-DC (both) DC
b. CHECKDeflection accuracy is within the limits
given in Table 5-3 for each CH 1 VOLTS/DIV switch set
ting and corresponding standard-amplitude signal. When at
the 20 mV VOLTS/DIV switch setting, rotate the CH 1
VOLTS/DIV Variable control fully counterclockwise and
CHECK that the display decreases to 2 divisions or less.
Then return the CH 1 VOLTS/DIV Variable control to the
CAL detent and continue with the 50 mV check.
Table 5-3
Deflection Accuracy Limits
VOLTS/DIV
Switch
Setting
Standard
Amplitude
Signal
Accuracy
Limits
(Divisions)
2 mV
10 mV
4.90 to 5.10
5 mV
20 mV
3.92 to 4.08
10 mV
50 mV
4.90 to 5.10
20 mV
0.1 V
4.90 to 5.10
50 mV
0.2 V
3.92 to 4.08
0.1 V
0.5 V
4.90 to 5.10
0.2 V
1 V
4.90 to 5.10
0.5 V
2 V
3.92 to 4.08
1 V
5 V
4.90 to 5.10
2 V
10 V
4.90 to 5.10
5 V
20 V
3.92 to 4.08
5-8

Table of Contents

Other manuals for Tektronix 2230

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix 2230 and is the answer not in the manual?

Tektronix 2230 Specifications

General IconGeneral
BrandTektronix
Model2230
CategoryTest Equipment
LanguageEnglish

Related product manuals