EasyManuals Logo

Tektronix 2230 Service Manual

Tektronix 2230
483 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #160 background imageLoading...
Page #160 background image
Adjustment Procedure—2230 Service
h. Set the SAVE/CONTINUE switch to SAVE (button
in).
i. Set the CH 2 VOLTS/DIV switch to 1 V.
j. CHECKThe display is compressed to 0.5 division
in amplitude.
k. Move the cable from the CH 2 OR Y input connector
to the CH 1 OR X input connector.
i. Set:
VERTICAL MODE CH 1
SAVE/CONTINUE CONTINUE (button out)
m. Repeat parts c through j.
n. Disconnect the test equipment from the instrument.
16. Adjust Attenuator Compensation
(C12, C11, C5, C4, C62, C61, C55, C54)
a. Set:
VOLTS/DIV (both) 0.1 V
STORE/NON STORE NON STORE (button out)
b. Connect the high-amplitude square wave output via
a 50 0 termination, a probe-tip-to-BNC adapter, and the
10X probe to the CH 1 OR X input connector.
c. Set the generator to produce a 1 kHz, 5-division
display and compensate the probe using the probe com
pensation adjustment (see the probe instruction manual).
d. Set the CH 1 VOLTS/DIV switch to 0.1 V.
e. Replace the probe and probe-tip-to-BNC adapter
with a 50 Q cable.
f. Set the generator to produce a 5-division display.
NOTE
Use Table 5-5 to identify the correct capacitor for
each channel adjustment.
g. ADJUST—The 10X Attn (Cl 2) for best front corner.
Table 5-5
Attenuator Compensation Adjustments
Adjustment
Channel 1
Channel 2
10X Attn (LF Comp)
Cl 2 C62
10X Attn (Input C)
C11
C61
100X Attn (LF Comp) C5
C55
100X Attn (Input C)
C4 C54
h. Replace the 50 Si cable and 50 Si termination with
the probe and probe-tip-to-BNC adapter.
i. Set the generator to produce a 5-division display.
j. ADJUSTThe 10X Attn (C11) for best flat top.
k. Repeat parts e through j until no further improve
ment is noted.
l. Set the CH 1 VOLTS/DIV switch to 1 V.
m. Replace the probe and probe-tip-to-BNC adapter
with the 50 S2 cable and 50 Si termination.
n. Set the generator to produce a 5-division display.
o. ADJUST—The 100X Attn (C5) for best front corner.
p. Replace the 50 S2 cable and 50 SI termination with
the probe and probe-tip-to-BNC adapter.
q. Set the generator to produce a 5-division display.
r. ADJUST—The 100X Attn (C4) for best flat top.
s. Repeat parts m through r until no further improve
ment is noted.
t. Set the VERTICAL MODE switch to CH 2.
5-12

Table of Contents

Other manuals for Tektronix 2230

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix 2230 and is the answer not in the manual?

Tektronix 2230 Specifications

General IconGeneral
BrandTektronix
Model2230
CategoryTest Equipment
LanguageEnglish

Related product manuals