4200A-SCS Parameter Analyzer User's Manual Section 2: Getting started
4200A-900-01 Rev. D/April 2020 2-7
To avoid overheating, operate the instrument only in an area with proper ventilation. Allow at
least eight inches of clearance at the back of the mainframe to assure sufficient airflow and
comply with the following guidelines:
Operate the instrument in a clean, dust-free environment.
Keep the fan vents and cooling vents from becoming blocked (sides and rear of the instrument).
Do not position any devices adjacent to the instrument that force air (heated or unheated) into
cooling vents. This additional airflow could compromise accuracy performance.
When rack mounting the instrument, ensure adequate airflow around the sides, bottom, and
back.
Do not rack-mount high power dissipation equipment adjacent to the 4200A-SCS.
To ensure proper cooling in rack environments with only convection cooling, place the hottest
equipment (for example, power supply) at the top of the rack. Place precision equipment, such
as the 4200A-SCS, as low as possible in the rack, where temperatures are the coolest. Adding
spacer panels below the instrument helps to ensure adequate airflow.
Basic system connections
This section provides basic system connections to get your 4200A-SCS set up and running. More
detailed connection information is in the "Connections and configuration" section of the
Model 4200A-SCS Reference Manual.
Connecting to protective earth
The 4200A-SCS must be connected to protective earth (safety ground) using the supplied
green-yellow ground cable. Failure to attach the ground wires to a known protective earth
may result in electric shock.