Connect the 4200A-SCS to the DUT
The hardware connections from the output terminals of the CVU to the MOS capacitor are shown in
the following figure. Connect one cable to each of the CVU terminals, then connect one SMA female
to BNC male adapter to the end of each cable.
The cables from the HCUR and HPOT terminals are coupled with the BNC tee adapter and are then
connected to the gate of the MOS capacitor. The cables from the LCUR and LPOT terminals are also
connected together with the BNC tee adapter and then are connected to the bulk of the MOS
capacitor.
In the Clarius application, you will configure HCUR to be the high terminal of the AC ammeter and
HPOT to be the high of the DC voltage source.
Figure 86: Connections from the 4210-CVU to the MOS capacitor