4200A-SCS Parameter Analyzer User's Manual Section 7: Make C-V measurements on a MOSCAP
4200A-900-01 Rev. D/April 2020 7-5
2. Select Projects.
3. Type MOScap into the search bar.
4. Select Search.
5. Select MOS Capacitor C-V project (cvu-moscap).
6. Select Create.
7. Select Yes when prompted to replace the existing project.
The project is displayed. This project has three C-V tests for a MOS capacitor, as shown in the next
figure.
Figure 88: MOS capacitor C-V project tree with three tests
Configure the test
To configure the C-V sweep test, you set the source and measure parameters, define the timing
parameters, and adjust the constants and formulas for extracting parameters from the C-V data.
This test makes a capacitance measurement at each step of a user-configured voltage sweep. A C-V
graph is generated from the acquired data and several device parameters are calculated in the
Formulator.
To configure the test:
1. Select Configure.
Figure 89: Configure highlighted