Section 7: Make C-V measurements on a MOSCAP 4200A-SCS Parameter Analyzer User's Manual
7-2 4200A-900-01 Rev. D/April 2020
Device connections
Using the supplied cables, make connections from the output terminals of the CVU to the MOS
capacitor.
Hazardous voltages may be present on all output and guard terminals. To prevent electrical
shock that could cause injury or death, never connect or disconnect from the 4200A-SCS
while the output is on.
To prevent electric shock, test connections must be configured such that the user cannot
come in contact with test leads, conductors, or any device under test (DUT) that is in contact
with the conductors. It is good practice to disconnect DUTs from the instrument before
powering up the instrument. Safe installation requires proper shields, barriers, and grounding
to prevent contact with test lead and conductors.