Section 5: Configure and use a Series 700 Switching System 4200A-SCS Parameter Analyzer User's Manual
5-4 4200A-900-01 Rev. D/April 2020
Connect the 4200A-SCS to the 7072
This section describes connections to the 7072.
To connect the 4200A-SCS and SMUs to the 7072:
Using four 4200-MTRX-X or 4200-TRX-X triaxial cables, make the following connections:
4200A-SCS GNDU FORCE to 7072 input terminal E
42x0 SMU channel 1 Force to 7072 input terminal A
42x0 SMU channel 2 Force to 7072 input terminal B
42x0 SMU channel 3 Force to 7072 input terminal C
To connect the 4210-CVU to the 7072:
1. Using the parts in the following figure, assemble a tee adapter to connect the 4210-CVU to the
7072.
Figure 53: 4210-CVU to 7072 adapter tee assembly