4200A-SCS Parameter Analyzer User's Manual Section 5: Configure and use a Series 700 Switching System
4200A-900-01 Rev. D/April 2020 5-5
2. Using four CA-447A SMA cables, make the following connections:
4210-CVU HCUR to adapter tee assembly 1
4210-CVU HPOT to adapter tee assembly 1
4210-CVU LPOT to adapter tee assembly 2
4210-CVU LCUR to adapter tee assembly 2
3. Connect adapter tee assembly 1 to input terminal G of the 7072.
4. Connect adapter tee assembly 2 to input terminal H of the 7072.
The connections are shown in the following figure.
Figure 54: 4200A-SCS to 7072 Matrix Card connections