Installation
6. Push OK Compensate Sig nal Paths from
the resulting side-bezel menu.
OK Com-
pensate
Signal
Paths
The calibration will take approximately
10 minutes to complete.
7. After calibration, verify that the status indicator
on the lower-bezel menu displays Pass.
Utility Page
Calibration
Signal Path
Pass
Factory
Pass
If it does not, then recalibrate the instrument
or have the instrument serviced by qualied
service personnel.
Service personnel use the factory calibration
functions to calibrate the internal voltage
references of the oscilloscope using external
sources. Refer t o your Tektronix eld ofce
or representative for assistance with factory
calibration.
NOTE. Signal Pa th Compensation does not include calibration to the probe tip. (See page 22, Compensating a Passive Volta ge
Probe.)
32 DPO3000 Series Oscilloscopes User Ma nual