EasyManuals Logo

Tektronix MSO70000C/DX User Manual

Tektronix MSO70000C/DX
1002 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #931 background imageLoading...
Page #931 background image
Oscilloscope R eference Sequence triggering (horizontal delay on)
Sequence triggering (horizontal delay on)
From the Trig menu, select A - B Trigger Sequence; then open the A -> B Seq tab.
Overview
Select the A-B Seq tab to use the A Event (main) trigger with the B Event trigger to capture complex data.
For information o n the controls, click the buttons.
Behavi
or
Select A - B Trigger Sequence to use the A Event (main) trigger with the B Event trigger to capture
compl
ex data. The A Event trigger arms the trigger system. The instrument then triggers on the B Event
trigger dened by events or time. You can also set up the instrument to delay the B Event Trigger by a
specied time period by turning Horizontal Delay Mode on.
When you click on the trigger window controls, the trigger graphic
(see page 386) shows the trigger
sequence.
What do you want to do next?
Le
arn a bout the A Event trigger types.
(see page 386)
Learn about B Event triggers. (see page 458)
Learn a bout Trigger Mode and Holdoff. (see page 474)
Return to the Trigger Setup control window. (see page 385)
DSA/DPO70000D, MSO/DPO/DSA70000C, DPO7000C, and MSO/DPO5000 Series 909

Table of Contents

Other manuals for Tektronix MSO70000C/DX

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MSO70000C/DX and is the answer not in the manual?

Tektronix MSO70000C/DX Specifications

General IconGeneral
BrandTektronix
ModelMSO70000C/DX
CategoryTest Equipment
LanguageEnglish

Related product manuals