Mask Test Summary
Overview
Use this window to view real time test results.
To Use
The Pass/Fail Summary is a read only field.
Behavior
The Pass/Fail Summary fields provide a real time summary of the test results. Waveforms or
Samples Tested shows how many waveforms or samples have been tested in relation to how many
were specified in the test setup.
Source indicates what signal source was used for the test.
Status indicates whether the test has passed, failed, or is passing. Note that passing appears in
yellow, passed in green, and failed in red.
Total Hits indicates the total number of data points that violated the mask.
Failed Wfms shows the total number of waveforms that failed during the test sequence.
Mask Hits Per Segment
Overview
Use this window to view the hits for each mask segment of the test.
To Use
The Hits per segment is a read only field.
Behavior
Each mask test can have up to 16 segments. The data point violations per segment are shown in
this window. Source indicates the input signal in use for the test. The segments are identified by
Seg followed by the number of the segment. Only the number of segments used in a test is shown.
A scroll bar appears if there are more segments used than can be seen at one time.

Table of Contents

Other manuals for Tektronix TDS5054B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS5054B and is the answer not in the manual?

Tektronix TDS5054B Specifications

General IconGeneral
BrandTektronix
ModelTDS5054B
CategoryTest Equipment
LanguageEnglish

Related product manuals