Chapter 7 – Register Operations
12
6
Modbus: Read/Write Register= 300, Len=1, 16-Bit Integer (lllll)
If type -= PRESET_TOTALIZER_RT, rrrrrppppp
Where:
rrrrr – Preset Totalizer Real Time Parameter Type as
custom_logic_preset_totalizer_real_time_prompt[]
{
{"GrossTotalizer" } // 00001
{"NetTotalizer" } // 00002
{"MassTotalizer" } // 00003
{"AuthGrossTot" } // 00004
{"AuthNetTot" } // 00005
{"AuthMassTot" } // 00006
{"UnauthGrossTot" } // 00007
{"UnauthNetTot" } // 00008
{"UnauthMassTot" } // 00009
{"RGrossTotalizer"} // 00010
{"RNetTotalizer" } // 00011
{"RMassTotalizer" } // 00012
{"RAuthGrossTot" } // 00013
{"RAuthNetTot" } // 00014
{"RAuthMassTot" } // 00015
{"RUnauthGrossTot"} // 00016
{"RUnauthNetTot" } // 00017
{"RUnauthMassTot" } // 00018
};
ppppp – Zero-based preset number (00000-00011).
Modbus: Read/Write Register=298, Len=1, 16-Bit Integer (rrrrr)
Modbus: Read/Write Register=299, Len=1, 16-Bit Integer (ppppp)
If type -= METER_TOTALIZER_RT, rrrrrppppplllll
Where:
rrrrr – Meter Totalizer Real Time Parameter Type as
custom_logic_meter_totalizer_real_time_prompt[]
{
{"GrossTotalizer" } // 00000
{"AuthGrossTot" } // 00001
{"UnauthGrossTot" } // 00002
{"CreepGrossTot" } // 00003
{"BckflwGrossTot" } // 00004
{"RGrossTotalizer"} // 00005
{"RAuthGrossTot" } // 00006
{"RUnauthGrossTot"} // 00007
{"RCreepGrossTot" } // 00008