Taking a Stat Test
Choose an area free of other nuclear gauges or radioactive
sources. Place the gauge on a level surface and ensure the
source rod is in the SAFE position.
To take a new stat test, press 1 at the Stat Test menu shown on
the previous page, and then press ENTER/START.
The gauge displays the progress of the stat test (twenty 1-
minute counts). Upon completion of the counts, the gauge
displays the results of the density stat test, including an
indication of whether the test passed or failed:
Press ENTER/START to view the results of the moisture stat
test. Follow the prompts to scroll through the results.
Reviewing the Stat Test
To review the results of the last stat test, press 2 at the Stat
Test menu shown on page 5–7. The gauge displays the results of
the density stat test as shown above.
Printing the Stat Test
To print the results of the last stat test to a serial printer or
computer via the gauge’s 9-pin serial port, press 3 at the Stat
Test menu shown on page 5–7. The gauge prompts the operator
to connect the printer to the gauge.
Connect the serial cable to the printer or computer serial port.
Press ENTER/START. The gauge prints or uploads the stat test
data and returns to the Stat Test menu.
NOTE
Gauge Serial Communications Cable is Part Number
113128.