Operation and Display 3-7
Particle
The amber light next to Particle flashes briefly each time a particle
is detected.
Flow
A green light comes on across from Flow when the flowrate is
5 ±0.5 cm
3
/s. A yellow light comes on above Flow when the flowrate
is too high; a yellow light comes on below Flow when the flowrate is
too low.
Auto
A green light comes on below Auto (next to Range) when you set the
display range to the auto-range mode.
Optics, Condenser, Saturator
A green light comes on across from Optics when the optics
temperature is ±2°C of the correct temperature. A green light comes
on across from Condenser and Saturator when the temperature of
each is within 0.5°C of their correct values. The microprocessor
normally controls the temperatures to within ±0.3°C of the correct
temperature for the condenser and saturator.
Laser
A green light comes on across from Laser when the laser is on and
operating at the correct power level.
Liquid Level
A green light comes on across from Liquid Level when the liquid
level sensor indicates that the reservoir is full.