9-6 Model 3022A Condensation Particle Counter
soapy water. Repair the leak and retest with the vacuum leak
test.
If you cannot locate and/or repair a leak yourself, contact TSI for
assistance.
HEPA Filter
Plug
Inlet
CPC
Compressed Air
10 psig
Figure 9-4
Checking for Leaks With Compressed Air
Installing an EPROM
Upgrading the EPROM is a special maintenance procedure. This
procedure is not considered routine and should only be attempted
by someone who is familiar with the operation of the CPC and who
is technically qualified.
You are encouraged to call TSI for assistance in performing special
maintenance. It may also be helpful to have the technician, tools,
and the CPC close to the telephone when discussing the problem
with a TSI technician. Refer to Chapter 9 for directions on
contacting a technical resource at TSI.