C–6 Series 3080 Electrostatic Classifiers
Read Temperature
RT (Read Temperature)
Returns the following temperatures:
Return: Sheath flow temperature [in
o
C],
Cabinet temperature [in
o
C]
Read User Settings
RUS (Read User Setting)
Returns values selected on front panel:
Return: Selected DMA voltage [in Volt]
Selected particle diameter [in nm]
Selected flow mode [D or S (D = Dual, S = Single)]
Selected electrical mobility [in cm
2
/(Vs)]
Selected sheath flow rate [in L/min]
Selected bypass flow rate [in L/min]
Voltage control mode [P or A (P= Panel, A= Analog)]
Read Measured Values
RMV (Read Measured Value)
Returns measured values:
Return: Selected Particle Diameter [in nm]
Measured DMA voltage [in Volt]
Measured sheath flow rate [in L/min]
Measured bypass flow rate [in L/min]
Measured absolute pressure [in mBar]
Measured sheath flow temperature [in
o
C]
Measured cabinet temperature [in
o
C]
Measured impactor flow rate [in L/min]
Selected electrical mobility [in cm
2
/(Vs)]
Voltage control mode [P or A (P= Panel, A= Analog)]
Selected flow mode [D or S (D = Dual, S = Single)]
Sheath flow status [0 or 1 (0 = not stable, 1 = stable)]
Bypass flow status [0 or 1 (0 = not stable, 1 = stable)]
High voltage status [0 or 1 (0 = error, 1 = OK)]
Measured impactor pressure drop [in cm H
2
O]
Selected DMA model [4-0]
Selected gas type [5-0]
Selected impactor [3-0]
Actual particle diameter [in nm]
Adjustable minimum particle diameter [in nm]
Adjustable maximum particle diameter [in nm]
Read Firmware Version
RFV (Read Firmware Version)
Returns the version of the firmware:
Return: Firmware version