Using Serial Data Commands C–7
Read Model Name
RMN (Read Model Name)
Returns the current model name and number as
displayed on the power on screen:
Return: “Electrostatic Classifier 3080”
Read On Time
ROT (Read On Time)
Returns the accumulated On Time hours of instrument,
sheath blower and bypass blower. (updated every
minute) stored in the EEPROM.
Return: Classifier On Time HH:MM
Sheath Blower On Time HH:MM
Bypass Blower On Time HH:MM
Read Flags
RFL (Read Flags)
Returns the status of the following flags:
Return: Sheath flow led status [0 or 1 (0 = not stable,
1 = stable)],
Bypass flow stable [0 or 1…],
High-voltage ok [0 or 1 (0 = not ok, 1 = ok)]
Definitions of Combined Set Commands
Set Particle Diameter
SPDx (Set Particle Diameter)
Set particle diameter:
SPD6.4 Set: particle diameter = 6.4 nm
SPD Return: particle diameter [in nm]
Set Voltage Mode
SVMx (Set Voltage Mode)
Set bypass flow rate:
SVMP Set: panel control
SVMA Set: analog control
SVM Return: P or A
Set High Voltage
SHVx (Set High Voltage)
Set DMA high-voltage:
SHV80.6 Set: dma voltage = 80.6 Volt
SHV Return: dma voltage [in Volt]