C–8 Series 3080 Electrostatic Classifiers
Set Flow Mode
SFMx (Set Flow Mode)
Set sheath blower mode:
SFMS Set: Single Blower
SFMD Set: dual blower
SFM Return: S or D
Set Sheath Flow Rate
SQSx (Set Sheath Flow Rate)
Set sheath flow rate:
SQS10.0 Set: sheath flow rate = 10.0 L/min
SQS Return: sheath flow rate [in L/min]
Set Bypass Flow Rate
SQBx (Set Bypass Flow Rate)
Set bypass flow rate:
SQB15.0 Set: bypass flow rate = 15.0 L/min
SQB Return: bypass flow rate [in L/min]
Set Flow Calibration Algorithm
SFCw (Set Flow Calibration Algorithm)
SFCS Set: Stepwise Linear Flow Calibration Algorithm
SFCP Set: Polynomial Flow Calibration Algorithm
SFC Return: P: Polynomial Flow Calibration Algorithm
S: Stepwise Linear Flow Calibration Algorithm
Set Flow Calibration Table
(Stepwise Linear Flow Calibration Algorithm)
SCTw,x,y,z (Set Calibration Table)
SCTw,x,y,z Set: w = 0: sheath flow calibration
x = row of calibration table (1 x 50)
y = raw sheath flow rate in (100 * L/min)
z = calibrated sheath flow rate in (100 *
L/min)
w = 1: bypass flow calibration
x = row of calibration table (1 x 50)
y = raw bypass pressure in (100 * mmH
2
O)
z = calibrated bypass flow rate in (100 *
L/min)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the TSI Instruments 3080 Series and is the answer not in the manual?

TSI Instruments 3080 Series Specifications

General IconGeneral
BrandTSI Instruments
Model3080 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals