C–8 Series 3080 Electrostatic Classifiers
Set Flow Mode
SFMx (Set Flow Mode)
Set sheath blower mode:
SFMS Set: Single Blower
SFMD Set: dual blower
SFM Return: S or D
Set Sheath Flow Rate
SQSx (Set Sheath Flow Rate)
Set sheath flow rate:
SQS10.0 Set: sheath flow rate = 10.0 L/min
SQS Return: sheath flow rate [in L/min]
Set Bypass Flow Rate
SQBx (Set Bypass Flow Rate)
Set bypass flow rate:
SQB15.0 Set: bypass flow rate = 15.0 L/min
SQB Return: bypass flow rate [in L/min]
Set Flow Calibration Algorithm
SFCw (Set Flow Calibration Algorithm)
SFCS Set: Stepwise Linear Flow Calibration Algorithm
SFCP Set: Polynomial Flow Calibration Algorithm
SFC Return: P: Polynomial Flow Calibration Algorithm
S: Stepwise Linear Flow Calibration Algorithm
Set Flow Calibration Table
(Stepwise Linear Flow Calibration Algorithm)
SCTw,x,y,z (Set Calibration Table)
SCTw,x,y,z Set: w = 0: sheath flow calibration
x = row of calibration table (1 ≤ x ≤ 50)
y = raw sheath flow rate in (100 * L/min)
z = calibrated sheath flow rate in (100 *
L/min)
w = 1: bypass flow calibration
x = row of calibration table (1 ≤ x ≤ 50)
y = raw bypass pressure in (100 * mmH
2
O)
z = calibrated bypass flow rate in (100 *
L/min)