Page 13
400V configuration
Property
Value: 5075
Total Power
6600W
(6*1100W)
Voltage
3Ph, 400VAC
Amperage
10A
Protection against electrical
shock
IEC 61010-1
Mains supply fluctuation
+/- 10%
3.4 Stand – by heating mode
The autoclave provides an option of heating the chamber in stand-by
mode between cycles with a very low power to reduce total cycle time
(1.6% of the total power only). The autoclave turns off automatically if
the
interval between the sterilization cycles is more than 2 hours.
3.5 Operating Conditions
This device is for indoor use only!
The sterilizer should be loaded only with autoclavable material!
The environment shall not exceed an ambient temperature of
40ºC and a relative humidity of 85% respectively.
• The operation altitude shall not be over 2000 meters (6561 feet)
(ambient pressure shall not be lower than 80 kPa (11.6 psi)).
• The autoclave shall not be used in a manner not specified in this
manual!
• Do not use the autoclave in the presence of dangerous gases.
• The packed or unpacked device shall be stored in ‘indoor’
conditions.
• Operate the autoclave only in the manner specified in the manual. If
the equipment is used in a manner not specified by the
manufacturer, the protection provided by the equipment may be
impaired.