Page 35
•
Sterilization: Sterilization temperature is maintained constant during
the sterilization time.
•
Cooling: N/A
•
Slow Exhaust: Steam is exhausted from the chamber at a slow rate,
until it reaches the pressure of 30 kPa above the ambient pressure.
Then steam
is exhausted out of the chamber at a fast rate until
pressure decreases to ambient pressure.
•
Drying: N/A
•
Ending: The system checks that End temperature and End pressure
have been reached, and then releases the door.
5.5 Program 5: Liquid A – Cooling (C – cooling models
only)
For liquids when the manufacturer recommends autoclaving at
temperatures of 121ºC (250F). for 15 minutes.
Cautions!
Both PT100 temperature sensors must be inside the bottles.
For proper sterilization, fill the bottles with approximately the
same amount of liquid.
Nominal parameters default settings
•
Sterilization temperature:121C (250F).
•
Sterilization time: 15 minutes.