Page 8
2 Safety Instructions
The autoclave has unique characteristics. Please read and understand
the operation instructions before first operation of the autoclave. The
following issues may require instructions guidance provided by the
manufacturer: how to operate the autoclave, the door safety
mechanism, the dangers involved in circumventing safety means, how
to ensure that the door is closed, and how to select a correct
sterilization program.
Make sure that you know where the main power switch is, where the
water cut-off valve is and where the steam and compressed air
disconnection valves are located.
Autoclave maintenance is crucial for the correct and efficient function of
the device. We enclose a log booklet that includes maintenance
recommendations, with every device.
The weekly spore test is part of the preventive maintenance plan, along
with the annual validation of the sterilization processes that ensures
appropriate temperature dispersion within the chamber.
Never use the autoclave to sterilize corrosive products, such as: acids,
bases and phenols, volatile compounds, or solutions such ethanol,
methanol or chloroform nor radioactive substances.
Never start using a new autoclave or a new steam generator, before
the safety, licensing and authorization department has approved it for
use.
All autoclave users must receive training in proper usage from an
experienced employee. Every new employee must undergo a training
period under an experienced employee.
A written procedure must be established for autoclave operation,
including daily safety tests, seal inspection and door hinge inspection,
smooth action of the closing mechanism, chamber cleaning, prevention
of clogging and preservation from corrosion, what is permitted and
what is prohibited for sterilization and choosing a sterilization program.
Liquids may be sterilized only with the “liquids” program. The container
must be covered but not sealed. Sealed bottles may only be sterilized
using a special program. The bottle must be either Pyrex or a
Borosilicate glass bottle.
When sterilizing plastic materials, make sure that the item can
withstand sterilization temperature. Plastic that melts in the chamber is
liable to cause a great deal of damage.
Individual glass bottles may be placed within an appropriate container
that will be placed on a tray. Never place glass bottles on the floor of
the autoclave. Never fill more than 2/3 of the bottle volume.
On closing the autoclave's door, make sure it is properly locked before
activating.