Page 77
9.4 Filling the Water Reservoir
Remove the water reservoir cover. Pour distilled water into the
reservoir through the opening on top of the autoclave until it reaches
the level of approximately 12 cm (0.5”0.75”) below the outlet of the
return tube. The water quantity is 1617 liters (4.24.5 US gal).
Use only water having the characteristics as per table in sec. 3.2. Tap
water may clog the system. A clogged system causes increase of
pressure, which prevent temperature from rising.
Caution!
Under no circumstance should water be filled above the level
defined above.