5.10 Program 10: Vacuum Test (PV, pre-vacuum models
only)
Vacuum Test is a test program with the following parameters:
Caution!
This is not a sterilization program!
Nominal parameters default settings
Vacuum pressure: 17.0 kPa
Vacuum time 1: 5 minutes
Vacuum time 2: 10 minutes
Operations sequence:
Vacuum is produced in the chamber down to P1=17 kPa. At this
stage all the valves close. The autoclave remains in this stage for 5
minutes. This period enables the condition in the chamber to reach
equilibrium.
After the 5 minutes have elapsed, the printer records the pressure
that is referred to as P2. At this point the test begins and lasts 10
minutes.
At the end of the test, the printer records the results. The pressure at
the end of the test is referred to as P3. The rate of change of P3-P2
shall not exceed 0.13 kPa/min.).