Page 43
5.11 Program 11: Warm-Up
Pre-warming the chamber at 80C (176F) without drying.
Caution!
This is not a sterilization program!
Nominal parameters default settings
•
Warm-up temperature: 80C (176F).
•
Warm-up time: 20 minutes.
Operations sequence:
•
Pulse low/Pulse high: N/A
•
Heating: Water enters the chamber and is heated by actuation of
electrical heaters until the warm-up temperature) is reached.
•
Warm up: Warm up temperature is maintained constant for the
warm-up time.
•
Sterilization: N/A.
•
Cooling: N/A.
• Fast Exhaust: the steam is exhausted out of the chamber at a fast
rate until pressure decreases to ambient pressure.
•
Drying: N/A.
Note: Some water may remain in the chamber.