Page 38
temperatures of 134C (273F) for 30 minutes. Drying stage is
available for PV (pre-vacuum) models only.
Nominal parameters default settings
•
Sterilization temperature: 134C (273F)
•
Sterilization time: 30 minutes
•
Drying time: 1 minute. (PV – pre-vacuum models only).
Operations sequence:
• (PV, pre-vacuum models only)
Pulse low/Pulse high: at one pulse it
will build vacuum down to 25 kPa. All exhaust from the chamber is
performed through the bio-hazard filter.
•
Water Inlet: Water enters the chamber.
•
Heating: The chamber is heated by actuation of electrical heaters
until the sterilization temperature is reached.
•
Sterilization: Sterilization temperature is maintained constant during
the sterilization time.
•
Cooling: N/A
•
Fast Exhaust: the steam is exhausted out of the chamber at a fast
rate until pressure decreases to ambient pressure.
Note: If the cycle fails, fast exhaust is performed through the bio-hazard
filter. Some water may remain in the chamber after a failed cycle. To
remove such water, press Start and repeat the cycle. If the problem
persists, call for service.
• (PV, pre-vacuum models only)
Drying: vacuum is built up and the
air pump is working.
•
Ending: The system checks that End temperature and End pressure
have been reached, and then releases the door.