40
4 SETTINGS AND CALIBRATIONS
40
SETTINGS AND CALIBRATIONS
Chapter
4
1. Excitation Voltage
Select an excitation voltage.
* If the excitation voltage is changed, perform re-calibration.
Operation
Setting range (2.5, 10V)
Main screenSetting First SettingY-axis SettingExc. Voltage
2. Unit
Select a unit.
Operation
Setting range (See "9-3.Unit setting list"on p.147.)
Main screenSetting First SettingY-axis SettingUnit
3. Zero Calibration
Register the present sensor input as zero.
Press with the load sensor unloaded.
Operation
Setting range (-3.333 to 3.333 mV/V)
Main screenSetting First SettingY-axis SettingZero Cal.
4. Equivalent Input Calibration
Input the rated output and rated capacity (display value)
described on the data sheet of the load sensor.
Also, set the decimal place here. Input the decimal point
together when inputting the display value.
Setting range
Rated output (-9.999 to 9.999 mV/V)
Operation
Display value (rated capacity) (-9999 to 9999)
Main screenSetting First SettingY-axis SettingNextEquiv. Cal.
5. Actual Load Calibration
Apply actual load to the load sensor, and input the load value at
that time.
Also, set the decimal place here. Input the decimal point
together when inputting the calibration value.
Setting range
Calibration value (-9999 to 9999)
Operation
Display value (display only)
Main screenSetting First SettingY-axis SettingNextActual Cal.