4 SETTINGS AND CALIBRATIONS
40
SETTINGS AND CALIBRATIONS
Chapter
4
1. Excitation Voltage
Select an excitation voltage.
* If the excitation voltage is changed, perform re-calibration.
Setting range (2.5, 10V)
Main screen→Setting→ First Setting→Y-axis Setting→Exc. Voltage
2. Unit
Select a unit.
Setting range (See "9-3.Unit setting list"on p.147.)
Main screen→Setting→ First Setting→Y-axis Setting→Unit
3. Zero Calibration
Register the present sensor input as zero.
Press with the load sensor unloaded.
Setting range (-3.333 to 3.333 mV/V)
Main screen→Setting→ First Setting→Y-axis Setting→Zero Cal.
4. Equivalent Input Calibration
Input the rated output and rated capacity (display value)
described on the data sheet of the load sensor.
Also, set the decimal place here. Input the decimal point
together when inputting the display value.
Setting range
Rated output (-9.999 to 9.999 mV/V)
Display value (rated capacity) (-9999 to 9999)
Main screen→Setting→ First Setting→Y-axis Setting→Next→Equiv. Cal.
5. Actual Load Calibration
Apply actual load to the load sensor, and input the load value at
that time.
Also, set the decimal place here. Input the decimal point
together when inputting the calibration value.
Setting range
Calibration value (-9999 to 9999)
Display value (display only)
Main screen→Setting→ First Setting→Y-axis Setting→Next→Actual Cal.