4 SETTINGS AND CALIBRATIONS
41
SETTINGS AND CALIBRATIONS
Chapter
4
6. Overload
Set the load at which the sensor is judged to be overloaded. The
Load OK output turns OFF when the load exceeds the set value.
Setting range (0 to 9999)
Main screen→Setting→ First Setting→Y-axis Setting→Next→Overload
7. Increment
Set the minimum unit of load display. Loads are displayed in the
set increments.
Setting range (1, 2, 5, 10)
Main screen→Setting→ First Setting→Y-axis Setting→Next→Increment
8. Analog Filter
The low-pass filter of the set cut-off frequency is inserted with
respect to the input of the load sensor.
Setting range (10, 30, 100, 300 Hz)
Main screen→Setting→ First Setting→Y-axis Setting→Next→Next→Analog Filter
9. Digital Filter
Loads are moving-averaged by the set number of times.
Setting range (0, 2 to 999 Times)
Main screen→Setting→ First Setting→Y-axis Setting→Next→Next→Digital filter
10.Digital Zero Limit
Set the limit of load that can be digital-zeroed. If Digital Zero is
performed with a load larger (smaller) than the zero-calibrated
load by the set load, an error will result, and the load will not be
zeroed.
Setting range (0 to 9999)
Main screen→Setting→ First Setting→Y-axis Setting→Next→Next→DZ Limit