50
4 SETTINGS AND CALIBRATIONS
50
SETTINGS AND CALIBRATIONS
Chapter
4
④Displacement
Measurement stops when the forced-stop or the
displacement reaches the Measurement Stop Level.
⑤Displacement Stop
Measurement stops when the forced-stop or the
displacement stops advancing to the time specified
under Measurement Stop Level.
4. Measurement Stop Level
Set load, time or displacement when the Measurement Stop
Condition is “Load,” “Time,” “Displacement” or “Displacement
Stop.”
Setting range
Load (-9999 to +9999)
Time (0.1 to 10.0 sec.)
Displacement (-9999 to +32000)
Operation
Displacement Stop (0.1 to 10.0 sec.)
Main screen→Setting→Work Setting→Start Setting→Stop Level
Load
Stop
displacement
Measurement
Section
Displace-
ment
Stop
level
Stop point Stop point
×
Not crossing
×
Not crossing
〇
Crossing in the
negative-to-
〇
Crossing in the
positive-to-
変位 変位 変位 変位
Stop
level
Stop
level
Stop
level
positive direction negative direction
Load
Displacement
not having
Displace-
ment
Measurement
Section
advanced for
the set time