Program Mode 5-5
Fill 2 Time
This parameter sets the time period to pressurize the test component
to the set test pressure. The programming
range is 0.0 to 5000.0
seconds and the default value is 0 seconds. If this parameter is set to
0 seconds, no valve or phase actions are performed.
Stabilize Time
This parameter sets the time period to allow the pressure to stabilize
before a leak rate measurement is performed. The programming
range is
0.3 to 5000.0
seconds and the default value is 0.3 seconds.
Test Time
This parameter the time period over which the leak rate is
determined. The programming range is
0.0 to 5000.0
seconds and
the default value is 1.0 seconds.
Vent Time
This parameter sets the time period during which the test component
is vented to atmosphere. The programming range is
0.0 to 5000.0
seconds
and the default value is 0.0 seconds. Note: Vent time will
automatically be extended until pressure is vented below 1 psi.
Comp/Cal Data
Submenu used to modify the measure setup result value.
Linking
Submenu used to setup links to additional programs to be performed.
Program Menu Parameters