7-10 Uson 628 Operator’s Manual
7.3.4 Reject Level
• In order to set a realistic Reject Level, the maximum admissible leak rate of the
test component must be determined. This can be obtained from component
specifications, trials, comparisons with previous test components or any other
reliable source of information.
Once the maximum admissible leak rate is known, the Reject Level (RL) can be
calculated using the following equation:
If the Tester configuration has not been modified by the user, the configuration
parameters will be set to their default values. These values are the minimum
settings available. It is probable that a minimum setting of the Reject Level will
cause the Tester to reject all test components.