Program Mode 5-7
5.3.4 Data Menu
The data menu parameters are:
Data Menu Parameters
Parameter Description
Password
Password entry screen.
Last Log
Displays last test result information.
Logs Review
Displays previous test result information.
Statistics
Displays statistic information for the selected program. The reset
options are password protected when enabled. The options are:
Reset:
Clears selected program counters.
Reset Statistics:
Clears all program counters.
Note:
Process Capability (CPK) formula.The upper and lower
process capability Cpu and Cpl formula with Upper Specification Limit
(USL) and Lower Specification Limit (LSL) is:
C
pl
= (Process Mean – LSL)/(3*Standard Deviation)
C
pu
= (USL – Process Mean)/(3*Standard Deviation)
The Cpk is the smallest value of the Cpl or Cpu denoted:
C
pk
= Min (C
pl
, C
pu
).
Export
Exports information to USB as a backup or for transferring to another
tester.The data is saved to location \uson\628\<serial number>\data.
Options:
System (all settings), Config (tester config), Diff sensor,
gauge sensor, statistics, datalog, setup, individual program or
programs.
Update:
Saves selected option to USB,
Import
Imports information from USB to restore setting or from another
tester. The options are:
File:
Selects the data to be imported.
Serial Number:
Specifies the tester serial number data to be
imported.
Update:
Reads selected data.
This option is password protected when enabled.
Reports
Outputs the settings in a text format to Serial port, or USB.
The options are:
Data:
Selects report data to be created.
Program:
Selects program number for the Program or Statistics
reports.
Output:
Selects the destination.
Update:
Outputs selected date.
The USB option saves the data to location \uson\628\< serial
number> \Reports.
Clear Data
This option allows the following information to be cleared and set to
the defaults setting: All, Setup, Datalog, Logs, Reports.
This option is password protected when enabled.