Program Mode 5-9
Reset Mode
This parameter defines the tester’s response to a test result. The
options available are:
None:
No acknowledgement is required.
Hold on Reject:
The operator must acknowledge a reject result has
been detected by pressing the Abort button to continue testing.
Hold on Pass:
The operator must acknowledge a Pass result has
been detected by pressing the Abort button to continue testing.
Inf.Fill:
A reject condition selects the Infinite Fill step to allow the
operator to search for leak or leaks in the rejected component.
Test:
Automatically restarts the test cycle. Note cycle will stop if an
error, reject or reset condition occurs.
Spool Check
This parameter enables/disables equal value spool position check.
Audit Output
This parameter defines the destination for the destination for the
result audit log: The options available are:
Off:
Disabled (default).
Serial:
To external device (i.e. printer) connected to Serial port.
USB:
To file on external USB memory device connected USB host
port.
File:
To file in internal memory which can be accessed using the
Export option.
Notes:
1. Files are created daily with the name: Resultsyyyymmdd.log.
2. The USB option saves data to \uson\628\<serial number>\audit
Serial
This option allows the serial port communication parameters to be
changed. Default Parameters are:
Baud rate:
9600.
Parity:
None.
Data bits:
8.
Control:
None.
Password
This option controls the password protection for parameters:
Control:
The options are Off (default) to disable and On to enable the
password option.
Entry:
To set unique password for the tester.
Timeout:
This parameter defines the elapsed time allowed with no
user activity before being logged out.
Language
This parameter selects the language in which the messages are
displayed. The options are:
English, Chinese Spanish.
Brightness
This parameter controls the display brightness level. The option
range is from
0 to100%
and the default setting is
50%
.
Backlight
Timeout
This parameter defines the duration the back light remains on after
no input activity.The option range is from
0 (disabled) to 60 minutes
and the default setting is
0.
Audio
This parameter controls the audio output. The options are:
Off
(default) or
On.
Set up Menu Parameters
Parameter Description