12
PERFORM A STRING TEST
25. Ensure that the string phantom has been properly set up according to the instructions in the procedure
Set Up & Maintain the String Phantom on page80.
26. Turn the console off, and then turn the console on. The console is now in configuration mode.
27. On the Main Configuration screen, press the PROG PROBE button.
CABLE
TEST
PROG
PROBE
MFG
TEST
SETUP
NOW RUN
THIS UNIT IS NOT SETUP
28. On the Program Probe screen, press the PROG button.
PRINTPROG SCAN
AA
29. On the Program screen, verify the probe serial number is correct.
If the probe serial number is not correct, press ID, and then enter the correct serial number.
30. On the Program screen, press SAVE. The console returns to the Program Probe screen.
SEL–+
A Probe Type = 12DOME_120
Phi Offset = 5
Phi HYST = 2
Gain Offset = 0
Phi Time = 2.000
ID: 01234
# SCANS
924
ID SAVE
31. Place the probe into the string phantom probe insert, and then align the probe label with the recess in the insert.
Ensure the probe fits snugly in the insert.
Label alignment