Verathon Service Partner Manual: Diagnostics & Troubleshooting
ISSUE LOCATION ORDER CORRECTIVE PROCEDURE PAGE
Scan result volume is
inaccurate
Probe
1
Complete the section Perform a String Test of the
Certify the System procedure
12
2 Replace the Probe Tailcone 70
3 Contact Verathon Customer Care
Scan result shows
irregular bladder
shape
Probe
1 Identify Irregular Aiming Icons 24
2
Complete the section Perform a String Test of the
Certify the System procedure
12
3 Replace the Probe Tailcone 70
4 Contact Verathon Customer Care
Probe makes
irregular buzzing or
knocking sounds
Probe Contact Verathon Customer Care
Connect probe
image appears when
probe is connected
Probe
1 Program the Tailcone 72
2 Replace the Probe Tailcone 70
Probe has no internal
motion
Probe
1 Replace the Probe Tailcone 70
2 Contact Verathon Customer Care
Probe is leaking oil Probe Contact Verathon Customer Care
*
If an error message occurs during testing and you are unable to reproduce it, continue service as normal. If
the error message is reproducible, replace the PCB assembly.
Observed failure typically occurs within the probe rather than the console.

Table of Contents

Other manuals for Verathon BladderScan BVI 3000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Verathon BladderScan BVI 3000 and is the answer not in the manual?

Verathon BladderScan BVI 3000 Specifications

General IconGeneral
BrandVerathon
ModelBladderScan BVI 3000
CategoryMedical Equipment
LanguageEnglish

Related product manuals