67
Verathon Service Partner Manual: Repair & Replacement
11. On the Main Configuration screen, press the RUN button. This temporarily puts the console in standard
operating mode.
CABLE
TEST
PROG
PROBE
MFG
TEST
SETUP
NOW RUN
THIS UNIT IS NOT SETUP
SET THE dATE
12. On the Main Menu screen, inspect the date. If the date is correct, continue to the next procedure.
If the date is incorrect, continue with this procedure.
13. Press the ALT button.
BVI 3000
SCAN ALT
VERATHON
PRINT NOTES
13:2705/22/14
14. On the Alternate Menu screen, press the DATE button.
NAME TIMEDATE
Copyright 2013 by
VERATHON Corporation
VER: 3.020 (D949)
Box ID: 00001965
PREF
SELF
TEST
15. On the Date Preset screen, press the  button to toggle between increasing or decreasing the month
number, and then press the MM button until the correct month is displayed.
Note: Only values from 01 to 12 are allowed.
05/22/14
MM
DD
YY

DONE

Table of Contents

Other manuals for Verathon BladderScan BVI 3000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Verathon BladderScan BVI 3000 and is the answer not in the manual?

Verathon BladderScan BVI 3000 Specifications

General IconGeneral
BrandVerathon
ModelBladderScan BVI 3000
CategoryMedical Equipment
LanguageEnglish

Related product manuals