Technical specification
VF1807090083 EN revision 17
3.
Technical specification
Hand-Foot Contamination
Monitor
(4 vertical hand detectors)
HF-50
HF-100
HF-150
HF-300
HF-350
Type
K1010
K1011
K1012
K1013
K1014
Hand-Foot Contamination
Monitor
(2 horizontal hand
detectors)
HF-50
HF-100
HF-150
HF-300
HF-350
Type
K1175
K1176
K1177
K1178
K1179
Only figures with tolerances or limits are guaranteed data; any other values are only informative.
Detector parameters are described in detail in a separate document “Hand-Foot Contamination Monitor -
Radiometric Parameters” which forms an integral part of the Instructions for Use.
Basic data
Dimensions (W x H x D) - variant
with horizontal hand detectors
(599 x 1438 x 732) mm
(23⅝ x 56½ 29 in)
Dimensions (W x H x D) - variant
with vertical hand detectors
(822 x 1438 x 825) mm
(32⅜ x 56½ 32½ in)
Weight
52 kg (114.6 lb)
Power supply
(100 ÷ 240) V AC, (45 ÷ 65) Hz
Maximum input power
50 W (typical current consumption of the instrument is 0.2 A / 230 V)
Electric protection
IP 20
Number of detectors
6 (7)
4 x SCD-286 (hand detectors), HF-4 series: 2 x SCD-286
2 x SCD-525 (foot detectors)
1x SFP-100 (frisker) optional accessories
Communication with PC
1 x Service USB port, 1 x USB port for data import/export,
1 x Ethernet 10/100
Result indications
Display
Colour touch-screen (320 x 240) pixels
Visual signalization unit
Indicator lamp 24 mm, minimum luminous intensity 900 mCd
Acoustic signalization unit
Voice output
Incoming pulse signalization unit for a frisker