Bit
Value
Description
0
0
Measurement of contamination of hands and feet
1
Measurement using the frisker
1
0
Not contaminated
1
Contaminated
2
0
-
1
Not decided
1
3 - 7
always 0
-
Tab. 9 The meanings of bits of the contamination measurement bit mask
Bit
Value
Description
0
0
Disabled detector
1
Enabled detector
1
0
Not contaminated
1
Contaminated
2
0
-
1
Not decided
2
3
0
-
1
Detector overload
4 - 7
always 0
-
Tab. 10 The meanings of bits of the contamination measurement bit mask for the given detector
6.5.2 Clear archive, clear archive of error
You can purge the whole measurement archive on the PLC storage drive using the Clear archive option.
Press the Archive the Clear archive option in the service menu. The archive purging screen
appears on the display (Fig. 35).
The archive is be purged by pressing the Yes button. Successful purging is be confirmed by a
message on the display.
You are returned to the service menu by pressing the red arrow in the upper part of the screen.
Follow the similar procedure for Clear error archive.
1
This condition means a calculation error - the algorithm was not able to evaluate the result and the instrument reports an error.