Annex C.
Minimum settable alarm level versus background count rate
Depending on the location of the instrument, the background may differ from the detectors. When it is required
to meet a specific MDA, it is necessary to adjust the maximum contamination measurement time Tmax suitably.
The chart below shows dependency between the background count rate (n
bg
) and the minimum settable alarm
level (n*
almin
) in cps for different measurement times T
max.
Fig. 62 Background-Time dependence of MDA
Note: Calculation conditions: Tbuf = 120 s, coefficient for type I and II error = 3