Page 134 of 148 300004969171/B
IEC 61000-4-11 0% U
T
; 1 Cycle
70% U
T
; 25/30
Cycles
Single Phase: at
0
o
0% U
T
; 250/300
0% U
T
; 1 Cycle
70% U
T
; 25/30
Cycles
Single Phase: at
0
o
0% U
T
; 250/300
operation during power
mains interruptions, it is
recommended that the
Volcano System Family be
powered from an
uninterruptible power
supply or a battery.
Power frequency
(50/60 Hz) magnetic
field
IEC 61000-4-8
3 A/m 3 A/m
Power frequency magnetic
fields should be at levels
typical for commercial or
hospital environments.
NOTE U
T
is the a.c. mains voltage prior to application of the test level.
RFID Transmitter Specifications
Frequency/Bandwidth Frequency Modulation Radiated Power
13.56 MHz ± 7 kHz 100%ASK 300mW