Page 146 of 148 300004969171/B
Figure 60: DICOM Settings-Storage Servers Dialog Box
Storage Servers–Network Settings
Host Name: This field contains the host name of the remote computer to which
the Volcano system will be sending DICOM images.
IP Address: This field contains the Internet Protocol (IP) address of the system
to which the Volcano system will be sending DICOM images.
Find: The Find button performs a Network PING operation to find the
configured remote computer. If a remote computer host name is entered and no
IP Address is used, the IP Address of the successfully resolved host name is
automatically populated in the IP Address field. If both the host name and IP
Address are entered, then the IP Address is used for the PING instead of Host
Name resolution.
Storage Servers–DICOM SCP Settings
AE Title: This field is used to configure the Application Entity Title field of the
remote DICOM Store SCP (Service Class Provider) application to which the
Volcano system will be sending images. The factory default is “STORE_SCP”.
Port Number: This field is used to configure the port number to which the
remote DICOM Store SCP application will be listening for messages from the
Volcano system. The factory default is 104.
Response Timeout: This field is used to configure the amount of time (in
seconds) that the system will wait for a response from the remote DICOM Store
SCP (Service Class Provider) server after an image has been sent. This timer is
used to terminate the DICOM association (network connection) if