Maintaining 600E Pump Components 63
Figure 4-3 Transducer Replacement
6. Disconnect the transducer connector on the interconnect board (Figure 4-4). The
interconnect board is on the left rear of the pump. The transducer connection is on
the lower front of the board.
Figure 4-4 Transducer Connection on Interconnect Board
7. Remove the transducer and spacer from behind the front panel.
Installing a New Transducer
1. Line up the spacer and transducer and replace the two screws on the front panel.