<4. Setting converter>
4-21
IM 12A01G01-03EN 1st Edition : Mar. 25, 2021-00
Table 4.5 Error list
Location Alarm Alarm name default*1
System
I/O Alarm Fail safe occur F
Maintenance status Outputs in HOLD 1/2/3/4 N
mA1/2/3/4 Output Maint. status N
Setting alarm
mA conguration error 1/2/3/4 F
Display 1/2/3/4 setting error C
Auto wash/cal. setting error C
Auto start date error C
Contact cong. error 1/2/3/4 F
pH
Device status Sensor communication error F
Measure alarm
Temperature too high S
Temperature too low S
pH too high S
pH too low S
Imped. (pH/ORP) too high M
Imped. (pH/ORP) too low M
Impedance (ref) too high M
Impedance (ref) too low M
SSA temp outside spec S
Sensor status
Sensor not detect F
Temp element not detect F
ID chip failure F
LE not detect F
Calibration Due N
imp1 glassbreak detect F
imp2 glassbreak detect F
SC
Device status Sensor communication error F
Measure alarm
Temperature too high S
Temperature too low S
Conductivity too high S
Conductivity too low S
Temp comp1 warning S
Polarization detected N
SSA temp outside spec S
Sensor status
Sensor not detect F
Temp element not detect F
ID chip failure F
Calibration Due N
TB
Device status Sensor communication error F
Measure alarm
Tubidity high S
Turbidity low S
Photodetector current error S
Light source failure S
Auto Zero cal. result error S
Auto cal. Unstablet S
Sensor status Humidity sensor failure S
CL Device status Sensor communication error F