17-3
IM DL850E-01EN
Wiring unit Σ measurement functions
The following 24 measurement functions are available.
UΣ (average voltage): UrmsΣ (rms value),* UmnΣ (rectified mean value calibrated to the rms value),* UdcΣ
(simple average), UacΣ (AC component)
IΣ (av
erage current): IrmsΣ (rms value),* ImnΣ (rectified mean value calibrated to the rms value),* IdcΣ (simple
average), IacΣ (AC component)
PΣ (total active power), SΣ (total apparent power), QΣ (total reactive power), λΣ (average power factor), φΣ
(average phase difference)
WPΣ (total integrated power), WP+Σ (total positive integrated power), WP-Σ (total negative integrated power),
qΣ (total integrated ampere-hour), q+Σ (positive total integrated ampere-hour), q-Σ (negative total integrated
ampere-hour),
WSΣ (total apparent energy), WQΣ (total reactive energy), ZΣ (average impedance), RSΣ (average series
resistance), XSΣ (average series reactance), RPΣ (average parallel resistance), XPΣ (average parallel
reactance)
* Y
ou can select either the rms value or the rectified mean value calibrated to the rms value (but not both). In
either case, the value is displayed as rms.
Delta math measurement functions
For details on line voltages and R, S, and T
points, see the wiring system
figure provided later.
3P3W→3V3A
The following 8 measurement functions are available.
Urs (R-S line voltage): Urms3 (rms value),* Umn3 (rectified mean value calibrated to the rms value),* Udc3
(simple average), Uac3 (AC component)
It (phase current): Irms3 (rms value),* Imn3 (rectified mean value calibrated to the rms value),* Idc3 (simple
average), Iac3 (AC component)
3V3A→3P4W
The following 13 measurement functions are available.
Ur (R-N voltage): Urms1
(rms value),* Umn1 (rectified mean value calibrated to the rms value),* Udc1 (simple
average), Uac1 (AC component)
Us (S-N voltage): Urms2
(rms value),* Umn2 (rectified mean value calibrated to the rms value),* Udc2 (simple
average), Uac2 (AC component)
Ut (T-N line voltage): Urms3 (rms value),* Umn3 (rectified mean value calibrated to the rms value),* Udc3
(simple average), Uac3 (AC component)
In (neutral line current)
3P4W→3V3A
The following 13 measurement functions are available.
Urs (R-S voltage): Ur
ms1
(rms value),* Umn1 (rectified mean value calibrated to the rms value),* Udc1 (simple
average), Uac1 (AC component)
Ust (S-T voltage): Urms
2 (rms value),* Umn2 (rectified mean value calibrated to the rms value),* Udc2 (simple
average), Uac2 (AC component)
Utr (T-R line voltage): Urms3 (rms value),* Umn3 (rectified mean value calibrated to the rms value),* Udc3
(simple average), Uac3 (AC component)
In (neutral line current)
* You can select either the rms value or the rectified mean value
calibrated to the rms value (but not both). In
either case, the value is displayed as rms.
Other measurement functions
The following 3 measurement functions are available.
η (efficiency): Motor ef
ficiency, power efficiency
Uubf (three-phase voltage unbalance factor)
Iubf (three-phase current unbalance factor)
17 Power Math

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA ScopeCorder DL850E and is the answer not in the manual?

YOKOGAWA ScopeCorder DL850E Specifications

General IconGeneral
BrandYOKOGAWA
ModelScopeCorder DL850E
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals