Version 8.0
G-30-1682-en Page 149
S88 / OPMI Lumera T Preparations for use
Cable connections for the 1Chip HD camera
Consult the Instructions for Use for the "1Chip HD Camera" (G-30-1946) for
details of the required cable connections.
You can find a list of the video accessories available for use with your surgical
microscope in the section "video accessories for the 1Chip HD camera option"
on
page 148.
CAUTION Risk of electric shock caused by unapproved or defective accessories!
R506
Unapproved or defective accessories can cause increased leakage current.
The user and patient may suffer an electric shock if the signal interfaces or
video outputs are touched.
• When configuring your system, comply with the requirements of
IEC 60601-1 (IEC 60601-1-1 is still valid in some countries).
CAUTION Risk of tripping over cables lying on the floor!
R507
If cables are installed inappropriately on the floor, this presents an increased
tripping hazard and interferes with the flow of work. By tripping and falling
down, users may suffer tissue trauma.
• If possible, cables should only be laid outside of the area of movement.
CCU
DVI
DC 12V - 15V
from the
suspension
system
1Chip HD camera on the beam splitter, Splitter 20
S-Video HD SDI
Camera cable
permanently installed
in suspension system
(max. 5m)