Preparations for use S88 / OPMI Lumera T
Version 8.0
Page 154 G-30-1682-en
Adjusting the downward travel limit
• Loosen the adjustment screw (1) by a few turns.
• Press one of the release buttons for the surgical microscope's magnetic
brakes and lower the surgical microscope until it can be focused on the
surgical field (depending of the width of the objective lens) while still
maintaining a safe distance between the microscope and the surgical
field.
• Turn the adjustment screw (1) clockwise as far as it will go.
• Lower the surgical microscope to its lowest position again, and check the
distance between the microscope and the surgical field.
CAUTION Risk of injury caused by lowering of the surgical microscope!
• Do not adjust the suspension arm when it is positioned above the patient.
• Limit the stroke of the suspension arm to prevent any contact with the
patient should the surgical microscope be lowered accidentally.