What to do in the event of malfunctions S88 / OPMI Lumera T
Version 8.0
Page 196 G-30-1682-en
Lamp failure in the Superlux Eye light source
The indicator lamp (5) is lit when the lamp has failed or if the lamp module is
defective. After activation and ignition of the backup lamp, the indicator lamp
turns off again.
CAUTION Risk of injury caused by lamp rupture!
Lamp rupture (audible as a loud bang) may lead to jamming of the lamp
module and/or failure of the electronics modules.
• Before opening the lamp housing, ensure that the device is moved to
a position in which possibly falling particles cannot put the patient or user
at risk.
• Do not continue using the system if the lamp module is jammed or the
illumination is no longer operational due to defective electronic modules.
Contact our Service Department.
NOTE Overheating of the lamp module!
If ventilation grids are covered e.g. by drapes, this may lead to overheating
of the ventilation modules and deactivation of the lamp.
• Never cover the ventilation grids.
• Lamps switched off due to overheating will be automatically switched on
again when they have cooled down.