Operation S88 / OPMI Lumera T
Version 8.0
Page 170 G-30-1682-en
When using a fundus viewing system (e.g. RESIGHT)
Only use accessories expressly certified by the manufacturer for combination
with the surgical microscope.
• If you have mounted the RESIGHT fundus viewing system on the underside
of the microscope and steeply tilt the microscope, the RESIGHT fundus
viewing system may unintentionally move in and injure the patient.
Remove the RESIGHT fundus viewing system before steeply tilting the
microscope.
CAUTION Risk of injury!
If the working distance of the accessories mounted on the underside of the
microscope is insufficient, the patient’s eye may be injured e.g. by unintentional
movement of the focus.
• Make sure that the range of movement is sufficient for the external focus
position when accessory equipment is attached.
• When using a fundus viewing system (e.g. RESIGHT 500 or RESIGHT 700
by ZEISS), which is usually placed between the surgical microscope and
the patient, make sure that the patient is neither put at risk nor injured
by the motorized focusing system or the movement of the suspension
system arm.