Safety measures S88 / OPMI Lumera T
Version 8.0
Page 24 G-30-1682-en
Stand safety equipment
1 Release bar
allows non-sterile persons to unlock the magnetic brakes of the
suspension system.
2 Adjustment screw for limiting downward travel
serves to set the minimum working distance from the surgical field in the
vertical direction. Check this setting
before each surgical procedure.
3 Locking knob
for the horizontal position of the suspension arm.
Before removing or attaching a module (microscope, tube, etc.) move the
suspension arm into its horizontal position. Pull out the locking knob and
turn it clockwise or counterclockwise through 180°. At the same time,
slightly move the suspension arm up and down until the lock snaps in.
Now the suspension cannot swing upward due to missing mass. After
mounting the module, unlock the suspension arm and perform the
balancing procedure.